analog_offset[4]
-1,0V
1,0V
Figure 2.8: For TRIGGER and GATE negative thresholds can be set to support negative signals
such as 16mA NIM.
TDC on Input TRIGGER
There is a TDC connected to the input TRIGGER. When used with the TDC, input TRIGGER
supports negative pulses only and analog offset[4] must be negative. The TDC creates packets
of type 8. These packets first contain a coarse timestamp and a payload that can be used to
calculate the trigger position with higher precision. The function ndigo process tdc packet() can
be used to replace the coarse timestamp with the precise timestamp. This function is described
in section 3.5 on page 39. TDC pulses must have a minimum duration of 3.3ns. The dead-time
of the TDC is 32ns.
2.3 Extension Card
This Option is only available for the Ndigo5G-10.
cronologic GmbH & Co. KG 8 Ndigo250M-14 User Guide